Electrical Requirements - AB Sciex 4000 Series Site Planning Manual

Table of Contents

Advertisement

Table A-3 Acquisition Computer
Equipment
Dell OptiPlex 9010
Monitor
Shipping weight
Table A-4 Other
Equipment
Bench (optional)
Line Adjustment
Transformer (optional)
Gas generator (optional)

Electrical Requirements

Return to
checklist.
WARNING! Electrical Shock Hazard: Use only qualified personnel for the installation
of all electrical supplies and fixtures, and make sure that all installations adhere to
local regulations and safety standards.
The total combined system (mass spectrometer and roughing pump) power consumption is 3200 VA (50 Hz or 60
Hz) at 230 VAC.
A Line Adjustment Transformer (PN WC04179) is required if the voltage is not in the recommended range (207
VAC to 242 VAC), but it is not less than 188 VAC or greater than 250 VAC.
Site Planning Guide
RUO-IDV-01-0563-A | D5040008 A
Height
Width
38.6 cm
57.4 cm (22.5 inches)
(15 inches)
46.36 cm
37.2 cm (15 inches)
(18.5 inches)
The acquisition computer and monitor are included in the Accessories box.
Height
Width
80 cm
150 cm (59 inches)
(32 inches)
20 cm (8 inches)
20 cm (8 inches)
Refer to the documentation for the gas generator system.
Site Requirements
Length
6.8 cm (3 inches)
18.2 cm (7.25 inches)
Length
Weight
75 cm (30 inches)
87 kg (192 lbs)
28 cm (11 inches)
15 kg (33 lbs)
AB SCIEX 4000 Series of Instruments
Weight
7.6 kg (16.8 lbs)
5.1 kg (11.25 lbs)
17 of 33