Item
Description
1
Branch circuit for the mass spectrometer. One outlet is required. It must be within 1.6 m
(63 inches) of the mass spectrometer.
2
Branch circuit for the roughing pump. One outlet is required. It must be within 1.6 m (63 inches)
of the roughing pump. The optional mass spectrometer bench can be connected to this branch
circuit.
3
(Optional) One branch circuit with one or more outlets for the optional gas generator. Contact
the manufacturer of the gas generator for more information. The optional mass spectrometer
bench can be connected to this branch circuit.
4
Additional branch circuit for the acquisition computer, monitor, and options, such as the optional
mass spectrometer bench.
Note: The optional mass spectrometer bench can be connected to branch circuits 2, 3, or
4.
International Requirements
Return to
checklist.
• For installations outside of North America, use locally approved standard connections and cables.
North American Requirements
Return to
checklist.
• Branch circuits for the mass spectrometer and roughing pump (branch circuits 1 and 2 in
page
18) must be 15 A 207 VAC to 242 VAC (typical 220 VAC), 50 Hz or 60 Hz. Receptacles are equipped
with CSA/NEMA 6-15R straight-blade receptacles. Refer to
• One branch circuit with one or more outlets might be required for the optional gas generator (branch circuit
3 in
Figure A-2 on page
receptacles required if a gas generator is used.
• The branch circuit for the acquisition computer and monitor (branch circuit 4 in
can be 15 A 120 VAC.
Site Planning Guide
RUO-IDV-01-0563-A | D5040008 A
18). Refer to the manufacturer s documentation for the branch circuits and
Figure A-3 on page
Figure A-2 on page
AB SCIEX 4000 Series of Instruments
Site Requirements
Figure A-2 on
20.
18)
19 of 33