Hamamatsu Photonics C8484-16C Instruction Manual page 31

Digital ccd camera
Table of Contents

Advertisement

(4) Sub-array readout
Readout of the sub-array repeats according to the sequence set internally.
Command
External control pulse :
Scan mode
Exposure time
Sub-array
Designated time
Minimum repeat cycle can be calculated using the formula below.
Minimum repeat cycle = readout time + external control pulse's LO interval
Exposure time can be calculated using the formula below.
Ex) When the external control pulse's LO interval {HI interval} is 40,
Exposure time = external trigger pulse LO interval {HI interval} - 2.2 μs
External trigger pulse
CCD exposure
Electrical shutter signal
Readout signal
from CCD chip
Digital data output
:
AMD E
(External control mode)
ATP N {ATP P}
:
SMD A
(Sub-array)
:
EMD L
(External setting)
:
SVO n / SVW m / SPX e
:
within 40 µs
{HI interval} (exposure time)
(MAX) (exposure lock-in time) + 152 μs (MAX) (readout
lock-in time)
C8484-16C Instruction manual Ver.1.5
Pulse width min. 10 µs
max. 2.2 µs
1×1 (SPX 1) 152 µs
2×2 (SPX 2) 152 µs
4×4 (SPX 4) 182 µs
Readout period