(3) 4×4 binning readout
Exposure and 44 binning readout repeat according to the sequence set internally.
Command
External control pulse :
Scan mode
Exposure time
Binning
1V (vertical)
Setting count
Minimum Repeat cycle can be calculated using the formula below.
Minimum repeat cycle = 28.0 ms (readout time) + set exposure time
Exposure time can be calculated using the formula below.
Exposure time = 38.45 µs + (n-1) × 103.6 µs
External trigger pulse
CCD exposure
Electrical shutter signal
Readout signal
from CCD chip
Digital data output
:
AMD E (External control mode)
ATP N {ATP P}
:
SMD S (Binning)
:
EMD E,EST n (Internal setting)
:
SPX 4 (4×4)
:
103.6 µs
:
1H to 12 930H (38.45 µs to 1.339 s)
C8484-16C Instruction manual Ver.1.5
(n = 1 to 12 930)
min. 1 µs
Exposure time = EST n
max. 2.2 µs
Readout period