Hamamatsu Photonics C8484-16C Instruction Manual page 30

Digital ccd camera
Table of Contents

Advertisement

C8484-16C Instruction manual Ver.1.5
(3) 4×4 binning readout
Exposure and 44 binning readout repeat according to the sequence set internally.
Command
External control pulse :
Scan mode
Exposure time
Binning
Designated time
Minimum repeat cycle can be calculated using the formula below.
Minimum repeat cycle = 28.0 ms (readout time) + external control pulse's
Exposure time can be calculated using the formula below.
Exposure time = external trigger pulse LO interval {HI interval} - 2.2 μs
Ex) When the external control pulse's LO interval {HI interval} is 40,
Exposure time = 40 μs - 2.2 μs (MAX) + 182 μs (MAX) = 219.8 μs (MAX)
External trigger pulse
CCD exposure
Electrical shutter signal
Readout signal
from CCD chip
Digital data output
28
:
AMD E (External control mode)
ATP N {ATP P}
:
SMD S (Binning)
:
EMD L (External setting)
:
SPX 4 (4×4 binning)
:
within 182 µs
LO interval {HI interval} (exposure time)
(MAX) (exposure lock-in time) + 182 μs (MAX) (readout
lock-in time)
Pulse width min. 10 µs
max. 2.2 µs
max. 182 µs
Readout period