AC40 Service Manual
Selecting EAR5A Insert phone
Select "Transducer Setup" (Figure 22)
Select "Insert phone" and select EAR5A
Enter the serial numbers for the left and right phones in the field "Left" and "Right"
Calibration of EAR5A
Equipment needed for calibrating the EAR5A:
Measuring amplifier
Acoustical IEC 60318-5 (IEC126) coupler
Frequency counter
Correction values for coupler
30° adaptor cable
Figure 22
Page 50