AC40 Service Manual

Checklist for calibration

Functional checks
Visual inspection, safety - Check power cord and accessory cords for
signs of wear or damage and general physical condition of
instrument.
Headband – Tension.
Cushions - Check condition; replace if needed.
Patch cords - Check plugs and leads for intermittency.
Patient response - Check for proper operation.
Pure-tone electroacoustic measurements
Frequency (Hz) - Actual measurement recorded (all frequencies).
Distortion (%) - Recorded as a percentage of THD on all frequencies.
Output levels (all transducers) - Recorded as a deviation from
standard
Linearity of all channels (all transducers) (dB) - Recorded as a
deviation from standard.
Rise/fall time (m/sec) - Actual measurement recorded.
Overshoot (dB) - Actual measurement recorded.
On/off ratio - Measured and recorded in dB down.
Crosstalk - Measured and recorded in dB down.
Pulse tones - Duration (m/sec) - Actual measurement recorded
(pass/fail).
SISI - Indicate pass or fail and deviation from standard.
Unwanted noise check - Record with (pass/fail):
Interrupter switch
Switch sounds
Mechanical sounds
Masking measurements
Narrow band noise output level - Measured and recorded as deviation
from standard, all transducers.
White noise output level - Measured and recorded as deviation from
standard, all transducers.
Speech electroacoustic measurements
Speech output level - Signal applied to input, VU meter at zero -
Measured and recorded as deviation from standard, all transducers.
Speech through external inputs - CD / tape / phono - VU meter at
zero - Measured and recorded as deviation from standard all
transducers.
Distortion - Recorded as percentage of THD.
Speech noise - Output measured and recorded as deviation from
standard.
Sound field measurements
Warble/FM output level - Actual measurement recorded (all
frequencies) as deviation from standard.
Speech inputs in sound field output level - Signal applied to input, VU
meter set at zero – Measured and recorded as deviation from
standard.
Narrow band noise output level - Measured and recorded as deviation
from standard, all transducers.
Speech/white noise output level - Measured and recorded as
deviation from standard, all transducers.
Distortion - Recorded as percentage of THD.
Operation of instrument should be checked for normal function post calibration.
Page 30