Swedish Speech Calibration - Interacoustics AC40 Service Manual

Clinical audiometer
Table of Contents

Advertisement

AC40 Service Manual
SWEDEN
Coupler IEC 60318-3 1998 (6ccm)
Speech Audiometer
Speech
STAF 1996
RETSPL MaxHL RETSPL MaxHL
Linear Type A
22.0
Equ. FF Type A-E
0.5
Not Linear
22.0
Speech WN
SWEDEN
Coupler IEC 60318-3 1998 (6ccm)
Speech Audiometer
Speech
IEC60645-2 1997
RETSPL MaxHL RETSPL MaxHL
Linear Type A
20.0
Equ. FF Type A-E
6.5
Not Linear
7.5
Speech WN
SWEDEN
Coupler IEC60318-1 with type 1 adaptor
Speech Audiometer
Speech
IEC60645-2 1997
Frequency
RETSPL MaxHL RETSPL MaxHL
Linear Type A
20.0
Equ. FF Type A-E
3.5
Not Linear
5.5
Speech WN
SWEDEN
Coupler IEC 60318-3 1998 (6ccm)
Speech Audiometer
Speech
STAF 1996
RETSPL MaxHL RETSPL MaxHL
Linear Type A
22.0
Equ. FF Type A-E
3.5
Not Linear
22.0
RETSPL
RETSPL
Speech
0.0
Speech Noise
0.0
Speech WN
2.5
TDH39
Speech Noise
STAF 1996
108
27.0
93
120
0.5
115
105
27.0
95
22.5
95
HDA 280
Speech Noise
IEC60645-2 1997
100
20.0
95
110
6.5
105
120
7.5
120
22.5
95
HDA 200
Speech Noise
IEC60645-2 1997
90
20.0
85
100
3.5
95
110
5.5
105
22.5
90
DD45
Speech Noise
STAF 1996
108
27.0
93
115
3.5
115
104
27.0
94
Sweden Free Field
ISO 389-7 2005
Binaural
45°
90°
RETSPL
RETSPL
-4.0
-5.5
2.0
-4.0
-5.5
2.0
-1.5
-3.0
4.5

Swedish Speech calibration

Speech WN
Speech
Speech WN
Speech
Speech WN
Speech
Speech WN
Speech
Speech WN
Free Field max HL is found by subtracting the
Monaural
Free Field Power
45°
90°
0° - 45° - 90°
RETSPL
RETSPL
Max SPL
-2.0
-3.5
-2.0
-3.5
0.5
-0.5
SWEDEN
EAR 3A
Coupler IEC 60318-5 2006
Speech Audiometer
Speech
Speech Noise
STAF 1996
STAF 1996
RETSPL MaxHL RETSPL MaxHL
21.0
99
26.0
22.5
SWEDEN
EAR 5A
Coupler IEC 60318-5 2006
Speech Audiometer
Speech
Speech Noise
STAF 1996
STAF 1996
RETSPL MaxHL RETSPL MaxHL
21.0
99
26.0
22.5
SWEDEN
CIR 22/CIR33
Coupler IEC 60318-5 2006
Speech Audiometer
Speech
Speech Noise
STAF 1996
STAF 1996
RETSPL MaxHL RETSPL MaxHL
21.0
89
26.0
22.5
SWEDEN
B71
Coupler 60318-6 2007
Speech Audiometer
Speech
Speech Noise
IEC60645-2 1997
IEC60645-2 1997
RETSPL MaxHL RETSPL MaxHL
55.0
60
55.0
57.5
Free Field max SPL
selected RETSPL value
Free Field Line
0° - 45° - 90°
Max SPL
90
100
85
100
87.5
97.5
Page 12
22.5
95
84
85
84
85
84
85
50
55