Selecting Dd45 Headset - Interacoustics AC40 Service Manual

Clinical audiometer
Table of Contents

Advertisement

AC40 Service Manual

Selecting DD45 headset

Select "Transducer Setup" (Figure 6)
Select "Headset phone" to select DD45 headset
Enter the serial numbers for the left and right phones in the field "Left" and "Right"
Calibration of DD45
Equipment needed for DD45 calibration:
Measuring amplifier
Acoustical IEC 60318-3 (IEC 303) coupler
Frequency counter
Correction values for coupler
Figure 6