Selecting Hda280 Headset - Interacoustics AC40 Service Manual

Clinical audiometer
Table of Contents

Advertisement

AC40 Service Manual

Selecting HDA280 headset

Select the "Transducer Setup" (Figure 14)
Select "Headset phone" and select HDA280 headset
Enter the serial numbers for the left and right phone in the field "Left" and "Right"
Calibration of HDA280
Equipment needed for the HDA280 calibration:
Measuring amplifier
Acoustical IEC IEC 60318-3 (IEC 303) coupler
Frequency counter
Correction values for coupler
Spacer for HDA280 to coupler IEC 60318-3 (IEC 303)
30° adaptor cable
Figure 14