Indications of the type plate
(example)
1,20 kW
5,3 A
230 V / 50Hz
230 V / 60 Hz
1 N ~
Explosion proof inner chamber
Ex classification acc. to 2014/34/EU
II -/3G c IIB T3-T1 Gc X
Max. temp. of inner surface 320 °C
Temp. class depending on setting of
temp. safety device
Max. operating pressure 125 +/- 25
mbar
Symbol
Information
CE conformity marking
Ex classification acc. to ATEX Directive 2014/34/EU, referring to the inner chamber
TÜV mark of conformity (Mark P)
Electrical and electronic equipment manufactured / placed on the market in the EC
after 13 August 2005 and to be disposed of in a separate collection according to di-
rective 2002/96/EC on waste electrical and electronic equipment (WEEE).
The equipment is certified according to Customs Union Technical Regulation (CU
TR) for Russia, Belarus and Kazakhstan
1.5
General safety instructions on installing and operating the vacuum drying
oven
With regard to operating the vacuum drying oven VDL and to the installation location, please observe the
DGUV guidelines 213-850 on safe working in laboratories (formerly BGI/GUV-I 850-0, BGR/GUV-R 120
or ZH 1/119, issued by the employers' liability insurance association) (for Germany).
BINDER GmbH is only responsible for the safety features of the chamber provided skilled electricians or
qualified personnel authorized by BINDER perform all maintenance and repair, and if components relat-
ing to chamber safety are replaced in the event of failure with original spare parts.
To operate the chamber, use only original BINDER accessories or accessories from third-party suppliers
authorized by BINDER. The user is responsible for any risk caused by using unauthorized accessories.
VDL (E2.1) 05/2016
Information
Nominal power
Nominal current
Nominal voltage ± 10% at the indicated power frequency
Current type
Explosion proof inner chamber
Ex classification according to ATEX Directive 2014/34/EU
Ex classification of the inner chamber:
II -/3G c IIB T3-T1 Gc X
Maximum temperature of the inner surfaces: 320 °C / 608°F
Temperature class acc. to IEC 60079-14 for the entire cham-
ber: T1, T2 or T3 depending on the setting of the temperature
safety device
Max. operating pressure (heating release): 125 +/- 25 mbar
page 9/108