8.1.3 General notes on programming temperature transitions

When exceeding the tolerance limits set in the user level (chap. 10), the program is interrupted until the
actual temperature value returns to within the tolerance range. During this program interruption, the LED
(7d) flashes. Therefore, the duration of the program may be extended due to the programming of toler-
ances
The programming is saved even in case of a power failure or after turning off the chamber.
Solvents and the resulting vapors can ignite at excessive drying temperatures.
Excess drying temperature.
Ignition and explosion hazard.
 Determine the maximum drying temperature in relation to the ignition temperature of
the solvent according to the information panel "Temperature setting" (chap. 2.5). Do
NOT exceed this temperature when presetting the set point on the RD3 temperature
controller.
 In the case of solvent mixtures, use the ignition temperature of the material with the
lowest ignition temperature.
 Before starting the drying process, set the temperature safety device (chap. 12) to the
permitted maximum drying temperature according to the information panel "Tempera-
ture setting" (chap. 2.5).
 The heating is only released when a vacuum of 125 ± 25 mbar / 3.69 ± 0.74 inHg) or
below is reached. Only then should you start the program.
After program rundown the controller returns to fixed value operation showing Normal Display and equili-
brates to the temperature value previously entered in fixed value entry mode.
Before starting the program, check the set-point value entered in fixed value entry mode.
After program rundown temperature will equilibrate to this value.
Deactivate the week program timer (factory setting, setting in the user level, chap. 10) before
starting a program.
8.2

Set-point entry for program operation

From Normal Display, press down button X/W for 5 sec to access the program editor. Then enter the set-
points one after the other in all program sections of a selected program.
You can enter two programs with up to 10 sections each or one program with up to 20 sections (setting in
the user level, chap. 10).
In order to avoid incorrect programming, we recommend entering the values of the program course into a
table (template in chap. 8.3).
Example of program table (with default setting "Ramp"):
Section
SEC
S01
S02
S03
S04
S05
VDL (E2.1) 05/2016
DANGER
Temperature set-point
[ °C]
TEMP
40
60
90
90
20
Section length
[hh.mm]
TIME
00:30
01:30
01:00
03:20
00:01
page 54/108