Vacuum drying ovens for flammable solvents
Hide thumbs Also See for vdl 23:
Table of Contents

Advertisement

High concentration of inert gas.
Risk of death by suffocation.
∅ Do NOT set up chambers in non-ventilated recesses.
 Ensure technical ventilation measures.
 Respect the relevant regulations for handling these gases.
 When decommissioning the vacuum drying oven, shut off the inert gas supply: Close
valve (6).
Install a pressure reducer for inert gas operation. Set the pressure reducer to a pressure slightly above
ambient pressure. Ensure that the pressure reducer will open. Do not change this setting in order to avoid
perturbation inside the oven and release of big quantities of inert gas after flooding the VDL.
Following evacuation, an inert gas, e.g., nitrogen, is led into the inner chamber via the fine dosing valve
for inert gas (6), until pressure compensation with the atmosphere occurs. Depending on the individual
application, you can perform a second evacuation and repeat the inert gas flooding.
When the inert gas fine dosing valve is open, a maximum of approx. 0.6 m³/h gas flows into the inner
chamber. The introduction of inert gas by under-flooding in the lower region of the inner chamber rear
wall and the extraction at the inner chamber ceiling allow an effective inert gas flushing.
15.

Options

15.1 ATEX connection kits for vacuum pumps VP4 or VP5 (option)

Connection kit for VP4 (Art. no. 8012-0621) consists of:
• Straining ring DN10/16 (3 pc.)
• Universal centering ring DN10/16 (3 pc.)
• Transition piece, adapter DN 16-10/8 (2 pc.)
• Vacuum hose 10/8, 2 m / 78.7 in
• Union nut M14 (2 pc.)
• Elbow DN 10/10
Connection kit for VP5 (Art. no. 8012-0622) consists of:
• Straining ring DN 10/16 (2 pc.)
• Straining ring DN 20/25
• Universal centering ring DN 10/16 (2 pc.)
• Outer centering ring DN 20/25
• Transition piece, adapter DN 16-10/8
• Transition piece, adapter DN 25-DN10/8 with union nut M14
• Union nut M14
• Vacuum hose 10/8, 2 m / 78.7 in
• Elbow DN 10/10
For mounting the VDL connection kits for VP4 or VP5, please refer to the mounting instruc-
tions Art. no. 7001-0152 supplied with each connection kit.