General Indications - Binder vdl 23 Operating Manual

Vacuum drying ovens for flammable solvents
Hide thumbs Also See for vdl 23:
Table of Contents

Advertisement

Display 1
Display 2
The program controller RD3 permits programming temperature cycles.
You can enter two programs with up to 10 sections each or one program with up to 20 sections (setting in
the user level, chap. 10).
When changing from 2 programs to 1 program or vice-versa, existing programs are deleted.
The maximum length of an individual program section can be set to either 99 h 59 min or to 999 h 59 min
(setting in the user level, chap. 10). This setting is then valid for all program sections.
Programming can be done directly through the keypad of the controller or graphically through the soft-
ware APT-COM™ 3 DataControlSystem (option, chap. 15.6) specially developed by BINDER.
5.2

General indications

The program controller RD3 offers several functional levels:
Normal Display / Fixed value operation:
• Display of the temperature actual value (display 1) and of the actual date and time (display 2).
• The oven is in fixed value operating mode, equilibrating to the entered set-points.
Fixed value entry mode (chap. 6)
• Entry of the temperature set-point for fixed value operating mode
• Entry of temperature set-points 1 and 2 for week program operation
Program editor (chap. 8)
• You can enter two programs with up to 10 sections each or one program with up to 20 sections (selec-
tion in the user level, chap. 10). Entry of temperature set-points in all program sections (chap. 8.1).
• Deleting a program section (chap. 8.4)
Program start level (chap. 9)
• Selection of an entered program
• Entry of settings affecting the program course, as "start delay time" or "number of program cycles"
• Program start
VDL (E2.1) 05/2016
Figure 14: RD3 program controller
(7a) LED Heating active
(7b) (no function)
(7c) (no function)
(7d)
LED lit: program operation
LED flashing: exceeding of the
tolerance limits in "Fixed value
entry mode" or in "Program op-
eration". In program operation:
program interruption.
page 37/108