Cleaning And Decontamination; Cleaning - Binder vdl 23 Operating Manual

Vacuum drying ovens for flammable solvents
Hide thumbs Also See for vdl 23:
Table of Contents

Advertisement

16.2 Cleaning and decontamination

Clean the chamber after each use to avoid potential corrosion damage by ingredients of the charging
material.

16.2.1 Cleaning

Disconnect the oven from the power supply before cleaning. Disconnect the power plug.
The interior of the chamber must be kept clean. Thoroughly remove any residues of the charg-
ing material.
Wipe the surfaces with a moistened towel. In addition, you can use the following cleaning agents:
Exterior surfaces,
inner chamber,
door gaskets
Expansion racks,
rack holders
Instrument panel
Filter of compressed air
supply
Zinc coated hinge parts
rear chamber wall
Do not use cleaning agents that may cause a hazard due to reaction with components of the device or
the charging material. If there is doubt regarding the suitability of cleaning products, please contact
BINDER service.
We recommend using the neutral cleaning agent Art. No. Art. Nr. 1002-0016 for a thorough
cleaning.
Any corrosive damage that may arise following use of other cleaning agents is excluded from
liability by BINDER GmbH.
Any corrosive damage caused by a lack of cleaning, is excluded from liability by BINDER
GmbH.
VDL (E2.1) 05/2016
Electrical hazard.
Danger of death.
∅ Do NOT spill water or cleaning agents over the inner and outer surfaces.
 Before cleaning, turn off the chamber at the main power switch and dis-
connect the power plug.
 Completely dry the appliance before turning it on again.
Standard commercial cleaning detergents free from acid or halides.
Alcohol-based solutions.
We recommend using the neutral cleaning agent Art. No. 1002-0016.
Standard commercial cleaning detergents free from acid or halides, no salt
solution or chlorinated solvents.
We recommend using the neutral cleaning agent Art. No. 1002-0016.
Standard commercial cleaning detergents free from acid or halides.
We recommend using the neutral cleaning agent Art. No. 1002-0016.
If necessary, use compressed air
Standard commercial cleaning detergents free from acid or halides.
Do NOT use a neutral cleaning agent on zinc coated surfaces.
DANGER
page 82/108

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Vdl 115Vdl 53

Table of Contents