Menu: Settings - Dräger Alcotest 9510 Technical Manual

Hide thumbs Also See for Alcotest 9510:
Table of Contents

Advertisement

14

Menu: Settings

14.1 "Service phonenumber" Function
This function allows the technician to enter the service phone
number associated with the instrument. When an instrument
disables itself it will automatically display a notification to the
operator about the change in status and instruct the operator
to take action and call the number entered in this function.
14.2 "Sound" Function
This function allows the technician to adjust the volume and
volume frequency of the 9510. Settings can be tested prior to
saving.
14.3 "Display" Function
This function allows the technician to calibrate/adjust display
features. The display can be rotated, tested for defective
pixels, and contrast and horizontal / vertical adjustment.
Dräger Alcotest 9510 Washington Technical Manual V3.1 2014
14.4 "Printer" Function
This function allows the technician to enable/disable the
internal and external printer and select the number of copies.
Printers can be tested prior to saving changes. The 9510 can
print externally to most printers (i.e. laser, inkjet).
14.5 "Database management" Function
This function displays the number of tests per test-type
currently stored in memory, with the option to download test
data to a USB stick by selecting "Copy to USB" or remove
stored data by selecting "Delete".
NOTICE
i
i
To ensure that authorized personnel are able to
generate reprints of recently performed test the
"Delete Older Measurement Tests" function should be
the primary method to delete data records from
instrument local memory.
Menu: Settings
23