Dräger Alcotest 9510 Technical Manual page 29

Hide thumbs Also See for Alcotest 9510:
Table of Contents

Advertisement

OPERATOR
49
TIMEOUT
50
PUMP ERROR 2
51
PRINTER ERROR
SMARTCARD PIN
52
ERROR
SMARTCARD
53
TIME ERROR
SMARTCARD
54
CANCEL ERROR
SMARTCARD
55
LOCK ERROR
IMCOMPLETE
56
TEST
Dräger Alcotest 9510 Washington Technical Manual V3.1 2014
X
X
X
Time for operator's
Perform another test and complete
response (e.g. no
the test sequence answering all
button pressed on
operator prompts without leaving
message box) is
the instrument idle for extended
expired
periods of time.
This error is most
Ensure ambient air is free of
likely triggered by a
alcohol vapor.
test with a very high
concentration of
alcohol and an issue
with alcohol in the
ambient air.
A hardware-related
Contact WSP technician
problem with the
internal printer.
Although the
internal printer is not
required in normal
X
circumstances for
Evidential Breath
Tests, it is a fault
that must be
addressed.
Contact WSP technician
Problem with
smartcard that can
X
cause problems with
communication with
the host PC.
Contact WSP technician
Problem with
smartcard that can
X
cause problems with
communication with
the host PC.
Contact WSP technician
Problem with
smartcard that can
X
cause problems with
communication with
the host PC.
Contact WSP technician
Problem with
smartcard that can
X
cause problems with
communication with
the host PC.
Status code is for information only
based on operator assessment that
the test should be terminated as an
incomplete test. No instrument
remedy required. Instrument will
offer operator opportunity to start a
Operator terminated
new test with the previously
test sequence with
entered data ret
Incomplete test
assessment.
Troubleshooting Table
29