Troubleshooting Table
EC AGING COMP.
61
ERROR
DIAGNOSTIC
CHECK FAILED
68
DRYGAS CHECK
ERROR
70
EC_SENSOR_ST
RESSED
71
OPERATOR_INPU
T_ERROR
72
OPERATOR_CER
T_EXPIRED
73
30
A problem was
observed with the
aging compensation
algorithm, caused
X
by an invalid time
difference.
Any of the internal
functionality tests
were outside of
tolerance
An empty or expired
drygas cylinder was
observed while
checking the
pressure of the dry
gas cylinder(s).
X
This error would
only be encountered
by a WSP
technician. This
error is only
applicable during a
calibration
procedure (like QAP
Calibration) and
may occur when the
EC sensor's alcohol
load is too high due
to previously
performed tests
During the data
entry sequence, the
operator entered an
invalid observation
start time (in the
X
future compared to
current instrument
time)
During the data
entry sequence, the
operator expiration
date associated with
X
the scanned
operator ID has
expired.
Dräger Alcotest 9510 Washington Technical Manual V3.1 2014
Contact WSP technician
Repeat test. If problem continues
contact WSP technician.
Contact WSP technician
Wait the time recommended by the
instrument.
Repeat test and either enter
correct observation start time or
begin new observation period.
Ensure an operator with a valid
expiration date (in the future) is
operating the breath test
instrument.