DELAYED START PROGRAM FUNCTION

When the run begins, the rotor will not accelerate past 3000 rpm until
the chamber pressure drops below 750 microns. At that time, the rotor
accelerates to set speed.
To vent the chamber, press
opened only when the vacuum has been completely vented and the
rotor is at rest.
Press
and
ENTER/RECALL

•UV7TV—

, the instrument will not be activated. If this occurs, press
•UV7TV—
and
ENTER/RECALL
After you have pressed
checks for a setting in the TIME field. If there is no entry, the instru-
ment beeps twice and the TIME display begins blinking. Enter a
value and press
ENTER/RECALL
again to begin the run.
As the rotor begins to accelerate, the green LED on the
begins to blink. When the rotor reaches set speed, the LED stops
blinking and remains on until the run ends or

•UVQR—

Use the
key to do the following:
•UVQR—
• Press
at any time to terminate the run. A green LED on the
key will blink to indicate that the rotor is decelerating; a tone will
sound when the rotor comes to a stop and the green LED will
turn off.
• Press
and
ENTER/RECALL
run conditions that existed when the rotor began decelerating.
These are called the STOP values.
• Press
,
ENTER/RECALL
accumulated ω
t at the time the rotor began decelerating. The ω
2
value is displayed in the SPEED display.
The delayed start program function allows you to delay the start of a
run so that it begins and ends at the time you require—without you
being there to start the run. For example, if you require a four-hour
run to end at 8:00 am, you can program the instrument to hold the
rotor at a specified temperature and time at zero speed and then auto-
matically begin the run at 4:00 am.
•X79WWI—
. The chamber door can be
•UV7TV—
to begin a run. If you press only¤
again to begin the run.
•UV7TV—
and
ENTER/RECALL
. Then press
ENTER/RECALL
•UVQR—
•UVQR—
within 5 seconds to display the
•UVQR—
ω
, and
in succession to display the
2
t
Run Preparation
, the instrument
•UV7TV—
and
•UV7TV—
key
is pressed.
2
t
2-9