Vega VEGABAR 54 Operating Instructions Manual page 41

Pressure transmitter
Hide thumbs Also See for VEGABAR 54:
Table of Contents

Advertisement

Carry out position correction
Carrying out min. adjustment
Carrying out max. adjustment
VEGABAR 54 - 4 ... 20 mA/HART
Setup with the indicating and adjustment module PLICSCOM
Proceed as follows:
1 Edit the offset value in the menu item "Position correction"
with [OK].
Position correction
Offset =
P
+0000.0 mbar
53.0 mbar
2 Set the requested offset value, e.g. the actual measured
value with reciprocal sign with [–>] and [+].
3 Confirm with [OK] and move to min. (zero) adjustment with
[–>].
Proceed as follows:
1 Edit in the menu item "Min. adjustment" the % value with
[OK].
Min. adjustment
+000.0 %
=
+0000.0 mbar
0000.0 mbar
2 Set the requested % value with [+] and [–>].
3 Edit the requested mbar value with [OK].
4 Set the requested mbar value with [+] and [–>].
5 Confirm with [+] and move to max. adjustment with [–>].
The min. adjustment is finished.
Information:
To adjust with a filling, you simply enter the indicated actual
measured value. If the adjustment ranges are exceeded, the
message is displayed "Outside parameter limits". The editing
procedure can be interrupted with [ESC] or the displayed limit
value can be accepted with [OK].
Proceed as follows:
1 Edit the % value in the menu item "Max. adjustment" with
[OK].
41

Advertisement

Table of Contents
loading

Table of Contents