Emerson X-STREAM Series Instruction Manual page 77

Non-incendive (ex na nc) configuration for zone 2 and division 2 installations gas analyzer
Table of Contents

Advertisement

Instruction Manual
HASXENE-IM-EX
12/2014
Signal
*)
The analog output is deactivated
None
A 0 or 4 mA signal is generated, e.g. to check the signal processing in an
0 mA
external system. Whether a 0 or 4 mA signal is generated, is set by the "Out
range" line (
A 20 mA signal is generated, e.g. to check the signal processing in an exter-
20 mA
nal system.
Comp1 ... 5
Gas concentration
Temp1 ... 5
Temperature
Press1 ... 5
Pressure
Flow1 ... 5
Flow
CalcA ... D
Result of calculator
RawVal1 ... 5
Raw value
RangeID1 ... 5 ID of selected range
*) Numbers 1 to 5 refer to components [chan-
nels] 1 to 5: In case of secondary parameters,
this means, the selected value is that of the
sensor assigned to the given component
(Press2 is the pressure value of the sensor
assigned to component 2).
Emerson Process Management GmbH & Co. OHG
3.9 Checking the Settings
Description
next page).
In contrast, capital letters A to D imply that
these calculator results are component
[channel] independent (Calc C is the result
of calculator C).
X-STREAM Non-Incendive
3-21