3.9 Checking the Settings
None
1013 hPa
1013 hPa
Good
X-STREAM Non-Incendive
Press
several times to return to SETUP
LEFT
and open MEASUREMENT.
Signal damping (set in DAMPING) allows
smoothing the output signal, but also affects
the response time of outputs and display:
• The factory setting is 0 seconds.
• The maximum possible t
the size of the internal sampling buffer and
the sampling rates of the installed measu-
ring principles/sensors.
• Multi-channel units: the value for each
channel must be entered separately.
The second page´s last line
enables the user
to enter the current ambient pressure ma-
nually, if no pressure sensor is installed,
or
• to view the current pressure, if a sensor
is installed (
INSTALLED OPTIONS).
If no pressure sensor is installed, enter the
current ambient pressure here and adjust
it, when signifi cant changes take place: this
improves the accuracy of the instrument.
time is limited by
90
3-19