Emerson X-STREAM Series Instruction Manual page 41

Non-incendive (ex na nc) configuration for zone 2 and division 2 installations gas analyzer
Table of Contents

Advertisement

Instruction Manual
HASXENE-IM-EX
12/2014
Enclosure purge option
The purge medium (e.g. to minimize CO
interference or for enhanced safety when
measuring corrosive or poisonous gases)
must be dry, clean and free of corro-
sives or components containing sol-
vents.
has to be free of components to be
measured, to minimize cross interfe-
rences.
Its temperature must correspond to the am-
bient temperature of the analyzer, but be at
least within the range 20...35 °C (68...95 °F).
Contact your Emerson Sales offi ce for infor-
mation on purge gas fl ow and pressure.
We recommend to always purge
the analyzer enclosure, if gases
are supplied, which may harm
analyzer components, if due to
a leak released into the analyzer
enclosure!
Emerson Process Management GmbH & Co. OHG
2.5 Installation - Gas Conditioning
2
X-STREAM Non-Incendive
Open reference option
In some cases, the measuring cell has an open
reference side, to be supplied with nitrogen.
This nitrogen
at least should be of quality 5.0, which
means nitrogen of purity ≥ 99.999 %.
If such gas is not available, the substitute
must be dry, clean and free of corrosi-
ves or components containing solvents.
has to be free of components to be
measured, to minimize cross interfe-
rences.
In any case, the gas temperature must cor-
respond to the ambient temperature of the
analyzer, but at least be within the range
20...35 °C (68...95 °F).
Pressure and gas fl ow must remain within
the values given in the
Specifi cations" section within this manual.
Perform a calibration each time
the source of this gas (e. g.
bottle) has changed!
„Measurement
2-7