X-STREAM Non-Incendive
The number of actually available signal out-
puts, and also the number of built-in modules
with screw-type terminals, varies according to
the confi guration of the unit.
Characteristics of terminals:
Accepted wire gauge:
Skinning length:
Thread:
Min. tightening torque:
All power and signal cables must end in a safe area or within a protective
enclosure (e.g. Ex e enclosure)!
Verify the power cord is layed with a distance of at least 1 cm (0.4 in) to any
signal cable to ensure proper insulation from signal circuits!
2-12
2.7 Installation - Electrical
0.14...1.5 mm
end sleeves not required
5 mm (0.2")
M2
0.25 Nm (2.21 in.lb)
EXPLOSION AND ELECTRICAL SHOCK HAZARD
A maximum of three modules with 36 termi-
nals each can be fi tted.
The terminals can be accessed by opening
the front panel of the instrument.
(26 AWG...16 AWG),
2
Emerson Process Management GmbH & Co. OHG
Instruction Manual
HASXENE-IM-EX
12/2014