Emerson X-STREAM Series Instruction Manual page 120

Non-incendive (ex na nc) configuration for zone 2 and division 2 installations gas analyzer
Table of Contents

Advertisement

X-STREAM Non-Incendive
A.7 Assignment of Plugs and Terminals
XSTA: Standard strip with standard and optional signals
Serial Interface
*)
Relay Outputs
**)
See table below
*)
**)
Configuration of relay output terminals as per stan-
dard factory setting (NAMUR status signals) . NAMUR
status signals are automatically configured Fail Safe!
Assignment of serial interface terminals
MOD 485/
MOD 485/
Terminal
2 wire
4 wire
P4.4
SER1
Common Common Common
P4.5
SER2
not used
not used
P4.6
SER3
not used
not used
P4.7
SER4
not used RXD1(+) not used
P4.8
SER5
TXD1(+) Common
D1(+)
P4.9
SER6
not used
not used
P4.10
7
not used
not used
P4.11
8
not used
RXD0(-)
P4.12
9
D0(-)
TXD0(-)
Signal Terminals Strips
A-18
A.7
7 Dig Inputs and 9 Dig Outputs each
Analog Outputs
Pin
9
8
7
6
RS 232
Pin no.
RS 232
1
Common
RXD
2
RXD
TXD
3
TXD
4
not used
5
Common
not used
6
not used
not used
7
not used
not used
8
not used
not used
9
not used
Service Port Connector -
Serial RS 232 interface
Assignment of Plugs and Terminals
XSTD: Optional strips with
Digital Outputs
Digital Inputs
1
/2
***)
st
nd
no. identifies inputs/outputs
on the 1
st
/2
nd
expansion board
Pin
5
4
Note!
3
Line and neutral terminals with built-in fuse holders
2
1
L= Line
N= Neutral
PE= Protective Earth
Power terminals
XSTI: Analog inputs
*) Alternatively on the curcuit card XASI,
the Jumper P2 can be plugged
**) Alternatively on the curcuit card XASI,
the Jumper P1 can be plugged
IN2
IN1
Mode
Mode
Analog Inputs
L
N
PE
Power cord
entry
Equipotential bonding
connector (thread:M5)
Emerson Process Management GmbH & Co. OHG
Instruction Manual
HASXENE-IM-EX
12/2014
Pin 1
Pin 8
Pin no.
Signal
1
TX+
2
TX-
3
RX+
6
RX-
other
not used
Ethernet connector
for Modbus

Advertisement