Emerson X-STREAM Series Instruction Manual page 37

Non-incendive (ex na nc) configuration for zone 2 and division 2 installations gas analyzer
Table of Contents

Advertisement

Instruction Manual
HASXENE-IM-EX
12/2014
Safe operation requires the instrument to be installed in a way not affecting
the ingress protection (IP66 / Type 4X)! Take care of the instructions for
installation.
Open covers or covers not properly closed affect the ingress protection and
result in unsafe operation! Do not operate the instrument without properly
closed covers!
Connecting and disconnecting non-incendive X-STREAM gas analyzers is
permitted only, if the instrument and all associated power & signal lines
are de-energized!
Failure to follow the instructions may cause explosion, property damage
and personal injury or death!
Do NOT operate damaged analyzers!
Take out of operation and take care for proper maintenance or repair!
Failure to follow may cause explosion, physical injury or death!
X-STREAM fi eld housings, intended for outside and wall mounted use, weigh
between 26 kg (57 lb) and 45 kg (99 lb) depending on options installed.
Two people and/or lifting equipment is required to lift and carry these
units.
Take care to use anchors and bolts specifi ed to be used for the weight of
the units!
Take care the wall or stand the unit is intended to be installed at is solid
and stable to support the weight!
Emerson Process Management GmbH & Co. OHG
2.2 General Safety Instructions
EXPLOSION HAZARD AND ELECTRICAL SHOCK HAZARD
EXPLOSION HAZARD BY DAMAGES
HEAVY INSTRUMENT
X-STREAM Non-Incendive
2-3