Stacking the Incubators
(continued)

Preliminary Cleaning

Installing Access Port
Installing Air Sample
Thermo Scientific
8. From the unit front, secure the base of the top unit to the exterior top
of the bottom unit by tightening the 5/16-18 stacking bolts using a ½"
(13mm) wrench or suitable tool. Refer to Figure 1-7.
9. Replace the base cover on the top unit.
1. Remove the protective plastic coating on the shelf supports, duct
sheets, and air duct, if present.
Caution Before using any cleaning or decontamination method except
those recommended by the manufacturer, users should check with the
manufacturer that the proposed method will not damage the equipment.
2. Using a suitable laboratory disinfectant, such as quaternary
ammonium, thoroughly clean all interior surfaces to remove any
residues which may remain from production of the incubator. Refer to
Section 5.
Locate the opening in the top left
corner of the interior chamber.
Filter
Remove the tape from the
opening on the outside of the
unit. Locate the stopper with filter
in the hardware bag. Install in the
opening inside the chamber
(Figure 1-9).
1. Remove the filter from the shipping bag.
Filter
2. Separate one section of the tubing from the filter. Install this section to
the fitting on the blower plate.
3. After installing the top duct, connect the filter assembly to the tubing
coming through the top duct.
4. Insert the free end of the air sample filter tubing into the larger hole in
the back of the blower scroll. See Figure 1-9 for completed
configuration.
Figure 1-9. Filter and Sensor Locations
Water Jacketed CO
Section 1
Installation and Start-Up
s
Incubator
1-7
2