Enabling Temp/Rh To Be Displayed; Enabling Co To Be Displayed - Thermo Scientific 3110 Operating And Maintenance Manual

3100 series forma series ii water jacket co2 incubator
Hide thumbs Also See for 3110:
Table of Contents

Advertisement

Enabling Temp/RH to
be Displayed
Enabling CO
/O
2
Displayed
Thermo Scientific
On units that are equipped with the RH option, the upper seven-segment
display on the control panel can be configured to display Temp
continuously, RH continuously, or toggle between Temp and RH. If the
units does not have RH, the upper display will always display temperature.
If temperature is set to ON and the RH is set OFF, temperature will be
displayed continuously. If temperature is set to OFF and RH is set to ON,
RH will be displayed continuously. If both are turned ON, the display
will toggle between the two. The factory setting will default to toggle
mode if the RH option is present.
1. Press the Mode key until the Config indicator lights.
2. Press the right arrow until DISP TMP XXX or DISP RH XXX is
displayed in the message center.
3. Press the up/down arrow to toggle the setting ON/OFF.
4. Press Enter to save the setting.
5. Press the Mode key to return to Run mode or right/left to go to
next/previous parameter.
to be
2
On models that are equipped with the O
display on the control panel can be configured to display CO
continuously, O
units does not have O
is set to ON and the O2 is set OFF, CO
If CO
is set to OFF and O
2
continuously. If both are turned ON, the display will toggle between the
two. The factory setting will default to toggle mode if the O
present.
1. Press the Mode key until the Config indicator lights.
2. Press the right arrow until DISP CO2 XXX or DISP O2 XXX is
displayed in the message center.
3. Press the up/down arrow to toggle the setting ON/OFF.
4. Press Enter to save the setpoint.
5. Press the Mode key to return to Run mode or right/left to go to
next/previous parameter.
continuously, or toggle between CO
2
, the lower display will always display CO
2
is set to ON, O
2
system, the lower seven-segment
2
2
and O
. If the
2
2
will be displayed continuously.
2
will be displayed
2
system is
2
Water Jacketed CO
Incubator
2
Section 3
Configuration
. If CO
2
2
3-9

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

312031113121313031313140 ... Show all

Table of Contents