Preventive Maintenance - Thermo Scientific 3110 Operating And Maintenance Manual

3100 series forma series ii water jacket co2 incubator
Hide thumbs Also See for 3110:
Table of Contents

Advertisement

Your equipment has been thoroughly tested and calibrated before shipment. Regular preventive maintenance is important to keep your unit
functioning properly. The operator should perform routine cleaning and maintenance on a regular basis. For maximum performance and efficiency, it
is recommended the unit be checked and calibrated periodically by a qualified service technician.
The following is a condensed list of preventive maintenance requirements. See the specified section of the operating manual for further details.
We have qualified service technicians, using NIST traceable instruments, available in many areas. For more information on Preventive Maintenance
or Extended Warranties, please contact Technical Services.
Cleaning and calibration adjustment intervals are dependent upon use, environmental conditions and accuracy required.
Tips for all incubators:
Do NOT use bleach or any disinfectant that has high chlorine content.
Use water with characteristics as described in Filling Water Jacket.
Avoid spraying cleaner on the CO
Do not use powdered gloves for tissue cultures.
Action
Check CO
and/or N
tank levels (tanks without optional gas guard monitor)
2
2
Inspect the inner and outer door latches and hinges for proper operation and excessive
mechanical wear; inspect the line cord insulation and inner door silicone gasket for
chips and cracks. Replace as required.
Check water level in humidity pan, ½" from top. See "Filling the Humidity Pan".
Check proper blower motor operation (air flow, noise, and vibration)
Verify and document CO
, humidity and temperature calibration, as applicable. See
2
Calibration section.
Verify O
calibration, as applicable. See Calibration section
2
Perform a complete decontamination procedure. Wipe down interior, shelves and side
ducts with disinfectant. Change or clean blower wheel and scroll. Clean top duct.
Clean humidity pan. Rinse everything well with sterile water. See "Cleaning the
Interior, "Cleaning the Humidity Pan".
Change HEPA and gas filters, if applicable (or as needed). See "Installing Access Port
Filter, "Installing Air Sample Filter", "Installing HEPA Filter".
Replenish rust inhibitor in water jacket. See "Adding/Replenishing Rust Inhibitor".
Thermo Scientific

PREVENTIVE MAINTENANCE

Incubators
sensor.
2
Daily
Weekly
Monthly
Before first use, after every service call and between experiments
to prevent cross contamination
More frequent decontamination may be required, depending on
use and environmental conditions
Water Jacketed CO
Preventive Maintenance
6
Yearly
2
Months
years
Incubator
4-5
2