Access Port Filter & Co Sensor Cover Plate; Installing The Air Sample Filter; Installing The Hepa Filter - Thermo Scientific 370 Operating And Maintenance Manual

370 series steri-cycle co2 incubator direct heat with sterilization cycle
Hide thumbs Also See for 370:
Table of Contents

Advertisement

Section 1
Installation and Start-Up
Access Port Filter &
CO
Sensor Cover Plate
2
Installing the Air
Sample Filter
Installing the HEPA
1-8
Steri-Cycle CO
Incubator
2
1. Locate the opening in the top left corner on the inside of the chamber.
Remove the tape covering the opening on the outside of the unit.
2. Locate the stopper with filter in the hardware bag. Lift the metal port
cover and install the assembly in the opening inside the chamber. See
Figure 1-12.
3. Also in the hardware bag is the CO
the thumbscrews. Refer to Figure 1-12 for the location.
Figure 12. Filter and Plate
1. Remove the filter from the shipping bag.
2. Install the air sample filter assembly to the black hose barb behind the
top duct.
3. Insert the other end of the filter assembly onto the metal tubing on the
top duct. Refer to Figure 1-13.
Caution Be careful when handling the filter. The media can be damaged if
it is mishandled. To avoid damage to the incubator, do not operate the unit
Filter
without the HEPA filter in place.
sensor cover plate. Install, using
2
Thermo Fisher Scientific

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

371380381370 series

Table of Contents