Teac GX-1 Instruction Manual page 96

Integrated recorder
Table of Contents

Advertisement

Settings
The following dialog box appears when you choose Setup and Params, and then
double-click the channel that you want to set.
Channel :
Select ENABLE to record with this channel, or select DISABLE not
to record with this channel. Enable both channels of a slot.
Sample Frequency : You can select 10 times the Base Sample Freq when you
turn on Multi Sampling check box.
Range :
Select the input mode (TTL/AC) and the frequency range.
Input Sensibility : Select the input sensitivity in the AC input mode. Only signals of
the set voltage, or higher, will be detected.
Pulse Counts Per Rotate : Enter the number of pulses in the range 1 to 255. This
setting will be reflected to SLOPE in the header file.
Channel Name : You can name the channel. Use up to 16 alphanumeric characters.
Moving Average : Select the number of samples for the moving average. (1 to 16)
All F/V Amp Same Condition : When this option is selected, the same parameters
are set for all signal conditioners of the same type.
Notes on Moving Average:
The sampling cycle for the moving average is 1 ms independently of the sampling
frequency setting. For example, when Moving Average is set to "5", the average
of five values—the values of 4 ms before, 3 ms before, 2 ms before, 1 ms before,
and the current instantaneous value—is recorded as the data at the moment.
When Moving Average is "5", the average of
these five values is recorded as the data at the
moment.
Instantaneous value at the
current sample timing
t (ms)
Amplifiers
4-29