Specifications
Number of channels :
Input format :
Input coupling :
Input resistance :
Gage ratio :
Excitation voltage :
Range :
Range accuracy* :
Temperature characteristics : 1%/10˚C
Balance adjustment method : Electronic (accuracy: ±1.2%)
Balance adjustment range :
Frequency response :
S/N ratio* :
Low pass filter :
A/D converter :
Sampling frequency :
Power consumption :
Input connector :
*At temperatures of +23˚C±5˚C
4-24
2
Balanced difference input
DC
1 MΩ
2.0
2 V DC, within 5 V ± 0.2%. Has remote sens-
ing.
±200, 500, 1000, 2000, 5000, 10000 µST
±2% (±200 to 1000 µST)
±1% (±2000 to 10000 µST)
The accuracy above is guaranteed when the
sensors are connected before turning on the
power of the GX-1.
±700% (each input range)
DC to 30 kHz (–3 dB)
65 dB (typical), 60 dB or better (range: 2000
µST; sampling: 5 kHz; LPF: 2.5 kHz)
25, 50, 250, 500, 2.5 k, 5 k, 24 kHz,
–3 dB at the cutoff frequencies
(–5 dB at 24 kHz)
Attenuation: –48 dB/oct, Butterworth
16-bit
1 Hz to 200 kHz
Approx. 3.4 W
HR25-7P-8P from Hirose Electric Co., Ltd.
Connect the cables from the sensors to the input con-
nectors before turning on the power of the GX-1.
Otherwise, the precision may be lowered.
Pin assignment for input connector
4
7
1
3
6
8
2
5
Pin
Signal
number
name
1
+BV
Excitation voltage
2
–IN
Signal input
3
–BV
Excitation voltage
4
+IN
Signal input
5
GND
6
+S
Remote sensing signal of
7
–S
excitation voltage
Notes:
• Connect +S to +BV, and –S to –BV at the sen-
sor when remote sensing is performed, or at the
input connector when not performed. Otherwise,
a correct excitation voltage cannot be obtained.
• Be sure to connect the shield of the sensor con-
nector and the cable shield.