6102 Micro-Bath User's Guide
Operating Parameters
9.10.2
9.10.3
40
info@Fluke-Direct.com
New Stir Speed setting
16
Press "SET" to accept the new Stir Speed.
The stir motor speed needs to be varied for best stability. Table 4 on page 28 shows
nominal settings for several fluids.
High Limit
The High Limit Parameter adjusts the upper set-point temperature. The factory default
and maximum temperature are set to 200°C. For safety, a user can ad just the High
Limit down so the maximum temperature set-point is restricted.
High Limit parameter
HL
Press "SET" to enable adjustment of the High Limit parameter.
Flashes "HL" and then displays the setting
HL
Current HL setting
H=126
Adjust the HL parameter using the "UP" or "DOWN" keys.
New HL setting
H=90
To store the new setting, press "SET". Press "EXIT" to continue without storing the
new value.
Cutout Reset Mode
The Cutout Reset Mode determines whether the cutout resets automatically when the
tank temperature drops to a safe value or must be manually reset by the operator.
Cutout Reset Mode parameter
CtorSt
Press "SET" to enable adjustment of the Cutout Reset Mode parameter. Nor mally, the
cutout is set for manual mode.
Flashes "CtorSt" and then displays the setting
CtorSt
Cutout set for manual reset
rst
To change to automatic reset mode press "UP" and then "SET".
Cutout set for automatic reset
Auto