Parts and Controls

The user should become familiar with the bath and its parts:

Back and Bottom Panel

Refer to Figure 1 on this page.
Power Cord – Underneath the calibrator is the removable power cord inlet that plugs
into an IEC grounded socket.
Power Switch – The power switch is located on the power entry module (PEM). The
PEM also houses the fuses and the dual voltage selector. The PEM and Heater Voltage
Switch (see below) allow the unit to be field switchable for 115 VAC (±10%) or 230
VAC (±10%) operation.
DISPLAY
HOLD
RS-232
POWER
115 VAC 3 A
230 VAC 1.6 A
50/60 Hz
115 V - 3
230 V -
HEATER
VOLTAGE SWITCH
Back View

Figure 1 6102 Back Panel and Bottom

Heater Voltage Switch – To be used only when changing the input voltage. (See
Section 8.1 for instructions on changing the input voltage.)
Caution: The input voltage and heater voltage switch settings should al ways be
the same value.
Serial Port – A DB-9 male connector is present for interfacing the calibrator to a
computer or terminal with serial RS-232 communications.
Fluke-Direct
A F
250 V
1.6 A F
250 V
Parts and Controls
Back and Bottom Panel
POWER
115 VAC 3 A
230 VAC 1.6 A
50/60 Hz
115 V - 3
A F
250 V
230 V -
1.6 A F
250 V
HEA TER
VOL T AGE SWITCH
HEATER
VOLTAGE SWITCH
Bottom View
.com
1.888.475.5235
17

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents