6102 Micro-Bath User's Guide
Serial Interface Parameters
9.11.3
9.11.4
42
info@Fluke-Direct.com
Adjust the value with "UP" or "DOWN" and then use "SET" to store the sam ple rate
to the displayed value. "EXIT" does not store the new value.
New sample period
SP= 60
Duplex Mode
The next parameter is the duplex mode. The duplex mode may be set to full du plex or
half duplex. With full duplex any commands received by the calibrator via the serial
interface are immediately echoed or transmitted back to the device of origin. With half
duplex the commands are executed but not ech oed. The duplex mode parameter is
indicated by,
Flashes "dUPL" and then displays the setting
dUPL
Current duplex mode setting
d=FULL
The mode may be changed using "UP" or "DOWN" and pressing "SET".
New duplex mode setting
d=HALF
Linefeed
The final parameter in the serial interface menu is the linefeed mode. This pa rameter
enables (on) or disables (off) transmission of a linefeed character (LF, ASCII 10) after
transmission of any carriage-return. The linefeed parameter is indicated by,
Flashes "LF" and then displays the setting
LF
Current linefeed setting
LF= On
The mode may be changed using "UP" or "DOWN" and pressing "SET".
New linefeed setting
LF= OFF