Fluke 3125 User Manual

Fluke 3125 User Manual

Detachable hot plate

Advertisement

Quick Links

Hart Scientific
3125
Detachable Hot Plate
Users' Manual
Rev. 622101

Advertisement

Table of Contents
loading

Summary of Contents for Fluke 3125

  • Page 1 Hart Scientific 3125 Detachable Hot Plate Users’ Manual Rev. 622101...
  • Page 2 Limited Warranty & Limitation of Liability Each product from Fluke's Hart Scientific Division ("Hart") is warranted to be free from defects in mate- rial and workmanship under normal use and service. The warranty period is one year for the Surface Cal- ibrator.
  • Page 3: Table Of Contents

    Table of Contents 1 Before You Start ......1 Symbols Used ......1 Safety Information .
  • Page 4: Before You Start

    1 Before You Start Symbols Used Before You Start Symbols Used Table 1 lists the International Electrical Symbols. Some or all of these symbols may be used on the instrument or in this manual. Table 1 International Electrical Symbols Symbol Description AC (Alternating Current) AC-DC...
  • Page 5: Safety Information

    2200 controller. • DO NOT connect the 2200 to a non-grounded, non-polarized outlet. • The 2200 Controller and the 3125 Detachable Hot Plate are a matched set. DO NOT mix and match sets. • HIGH VOLTAGE is used in the operation of this equipment. SEVERE...
  • Page 6 1 Before You Start Safety Information cautions. Before working inside the equipment, turn the power off and disconnect the power cord. • Always replace the fuse with one of the same rating, voltage, and type. • This instrument is intended for indoor use only. •...
  • Page 7: Cautions

    Wait until the power has stabilized before re-energizing the in- strument. Authorized Service Centers Please contact one of the following authorized Service Centers to coordinate service on your Hart product: Fluke Corporation, Hart Scientific Division 799 E. Utah Valley Drive American Fork, UT 84003-9775 Phone: +1.801.763.1600 Telefax: +1.801.763.1010 E-mail: support@hartscientific.com...
  • Page 8 22 Jianguomenwai Dajie Chao Yang District Beijing 100004, PRC CHINA Phone: +86-10-6-512-3436 Telefax: +86-10-6-512-3437 E-mail: xingye.han@fluke.com.cn Fluke South East Asia Pte Ltd. Fluke ASEAN Regional Office Service Center 60 Alexandra Terrace #03-16 The Comtech (Lobby D) 118502 SINGAPORE Phone: +65 6799-5588 Telefax: +65 6799-5588 E-mail: antng@singa.fluke.com...
  • Page 9: Introduction

    Hart Scientific’s Model 3125 Detachable Hot Plate is controlled by Hart Scien- tific’s 2200 controller and uses a precision platinum RTD as a sensor with a heater to control the temperature. The 3125 and 2200 are a matched set. The controller display shows the temperature and also the set-point tempera- ture.
  • Page 10: Specifications And Environmental Conditions

    Specifications and Environmental Conditions Specifications Temperature Range 35°C to 400°C (95°F to 752°F) Display Accuracy ±0.5°C to 200°C ±1.0°C to 400°C Stability ±0.2°C to 300°C ±0.3°C to 400°C Resolution 0.01° Uniformity ±0.3°C at 100°C ±0.6°C at 200°C ±0.9°C at 300°C ±1.4°C at 400°C Heating Time 25°C to 400°C: 22 minutes...
  • Page 11 • ambient relative humidity: maximum 80% for temperature <31°C, de- creasing linearly to 50% at 40°C • pressure: 75kPa-106kPa • mains voltage within ±10% of nominal • vibrations in the calibration environment should be minimized • altitude does not effect the performance or safety of the unit...
  • Page 12: Troubleshooting

    4 Troubleshooting Troubleshooting Troubleshooting Problems, Possible Causes, and Solutions In the event that the instrument appears to function abnormally, this section may help to find and solve the problem. Several possible problem conditions are described along with likely causes and solutions. If a problem arises, please read this section carefully and attempt to understand and solve the problem.
  • Page 13: Comments

    3125 Detachable Hotplate User’s Guide Problem Possible Causes and Solutions The display shows any Controller problem. The error messages signify the following problems with of the following: err 1, the controller. err 2, err 3, err Err 1 - a RAM error...

Table of Contents