6102 Micro-Bath User's Guide

Scan

9.3
9.3.1
9.3.2
34
info@Fluke-Direct.com
New units selected
Un= F
Scan
The scan rate can be set and enabled so that when the set-point is changed the bath
heats or cools at a specified rate (degrees per minute) until it reaches the new set-point.
With the scan disabled the bath heats or cools at the maximum possible rate.

Scan Control

The scan is controlled with the scan on/off function that appears in the main menu
after the set-point function.
Scan function off
Sc=OFF
Press "UP" or "DOWN" to toggle the scan on or off.
Scan function on
Sc=On
Press "SET" to accept the present setting and continue.
Accept scan setting

Scan Rate

The next function in the main menu is the scan rate. The scan rate can be set from
.1 to 99.9°C/min. The maximum scan rate however is actually limited by the natural
heating or cooling rate of the instrument. This is often less than 100°C/min, especially
when cooling.
The scan rate function appears in the main menu after the scan control function. The
scan rate units are in degrees C per minute.
Scan rate in°C/min
Sr= 10.0
Press "UP" or "DOWN" to change the scan rate.
New scan rate
Sr= 2.0
Press "SET" to accept the new scan rate and continue.