3
3.1 Specifications

3.2 Environmental Conditions

info@Fluke-Direct.com

Specifications and Environmental Conditions

The 6102 may not meet the specification published if a different fluid is used other
than distilled water or Dow Corning (200.10, 200.20) oil.
Do not exceed the boiling point or flash point temperatures of the fluid. To ensure that
the boiling and flash point temperatures are not exceeded adjust the High Limit and
Cutout parameters. To set these parameters, refer to Section 9, Controller Operation.
Table 2
Specifications
Range
Accuracy
Stability
Uniformity
Resolution
Operating Temperature
Heating Time
Cooling Time
Well Size
Size
Weight
Power
System Fuse
Safety
Although the instrument has been designed for optimum durability and trou ble-free
operation, it must be handled with care. The instrument should not be operated in an
excessively dusty or dirty environment. Maintenance and clean ing recommendations
can be found in the Maintenance Section of this manual.
The instrument operates safely under the following conditions:
ambient temperature range: 5 – 45°C (41 – 113°F)
ambient relative humidity: maximum 80% for temperature <31°C, de creasing
linearly to 50% at 40°C
mains voltage within ± 10% of nominal
vibrations in the calibration environment should be minimized
altitudes less than 2,000 meters
Fluke-Direct
Specifications and Environmental Conditions
35 to 200°C (95 to 392°F)
±0.25°C
±0.02°C at 100°C
±0.03°C at 200°C
±0.02°C
0.01°C/F
5 to 45°C (41 to 113°F)
25°C to 200°C: 40 minutes
200°C to 100°C: 35 minutes
1.9" dia. x 5.5" deep (48 x 139 mm)
5.5" W x 10.38" H x 8" D (14 x 26 x 20 cm)
10 lb. (4.5 kg) with fluid
115 VAC (±10%), 12.3 A, or 230 VAC (±10%), 1.1 A, switchable, 50/60 Hz,
270 W
115 V: 250 V, 3A, F
230 V: 250 V, 1.6A, F
Overvoltage (Installation) Category II, Pollution Degree 2 per IEC 61010-01