:FREQuency?

Parameters
Comments
104 Multimeter Command Reference
MEASure:FREQuency? [<
resolution
[,<
>|MIN|MAX|DEF]]
for all inputs between 3Hz and 300kHz.
Parameter Name
Parameter Type
range
<
>
resolution
<
>
The frequency function uses one "range" for all inputs between 3Hz and
300kHz. Querying the range will always return "3E+00". A frequency
measurement returns "
Range and resolution settings are listed below for the
parameters and after a module reset (
AUTO
PARAMETER
MIN
MAX
and module
DEF|AUTO
reset (
*RST)
range
>|MIN|MAX|DEF|AUTO
selects the frequency function and uses one range
Range of Values
numeric
numeric
3E-04 | 3E-05 | 3E-06
" if no input is applied.
0
RANGE
3E+00
3E+00
3E+00
Default Units
3E+00
,
,
MIN
MAX
DEF
).
*RST
RESOLUTION
3E+06
3E+04
3E+05
Hz
Hz
and
Chapter 3