Db:reference; Dbm:reference - Agilent Technologies Agilent E1412A User Manual

6 1/2 digit multimeter
Table of Contents

Advertisement

:DB:REFerence

Example
:DB:REFerence?
Example

:DBM:REFerence

Example
:DBM:REFerence?
Example
Chapter 3
CALCulate:DB:REFerence <
Relative Register. You must turn on the math operation e.g., execute
CALC:STAT ON before writing to the math register. You can set the relative value
to any number between 200dBm (the
stored in volatile memory.
Set the
DB
Reference Value
CALC:STAT ON
CALC:DB:REF 60
CALC:FUNC DB
CALCulate:DB:REFerence? [MIN|MAX]
Query the
Reference Value Set for the
DB
CALC:DB:REF?
CALCulate:DBM:REFerence <
Choose from:
,
,
50
75
93
,
,
, or
900
1000
1200
8000
math operation e.g., execute CALC:STAT ON before writing to the math register.
The dBm reference is stored in non-volatile memory.
Set the
DBM
Reference Value
CALC:STAT ON
CALC:DBM:REF 135
CALC:FUNC DBM
CALCulate:DBM:REFerence? [MIN|MAX]
Query the
Reference Value Set for the
DBM
CALC:DBM:REF?
value
>|MIN|MAX
MIN
and
Turn on the math operation.
Sets DB reference to 60 dBm.
Select the DB math operation. You can select
the calculate function at any time before or
after enabling the calculate state.
queries dB reference value.
DB
Query the DB reference value.
value
>|MIN|MAX
,
,
,
,
,
110
124
125
135
150
ohms.
= 50.
MIN
MAX
Turn on the math operation.
Sets DBM reference value to 135.
Select the DBM math operation. You can select
the calculate function at any time before or
after enabling the calculate state.
queries the dBm reference.
DBM
Query the DBM reference value.
Multimeter Command Reference
stores a relative value in the dB
MAX
values). The dB reference is
Math Operation
selects the dBm reference value.
,
,
,
,
(default),
250
300
500
600
= 8000. You must turn on the
Math Operation
,
800
75

Advertisement