Example

:STRing?

Example

:VALue

Parameters
Comment
Example
:VALue?
Example
82 Multimeter Command Reference
Enter Calibration Information to Record the Next Calibration Date
CAL:STR 'Cal 4/4/YY, Due 10/4/YY' Enter a calibration message to record the cal
CALibration:STRing?
(or a null string " " if nothing is present).
Query the Calibration Message
CAL:STR?
enter statement
CALibration:VALue <
cal_value
signal used by the calibration procedure. See the E1412A Service Manual, Chapter
5 "Adjustments", for a more detailed description of the multimeter's
calibration/adjustment procedures.
Parameter Name
cal_value
<
>
*RST
does not affect the calibration value.
Enter the Known Value for the Calibration Source Signal
CAL:VAL 10.0
queries the present calibration value.
CALibration:VALue?
Query the Calibration Value
CAL:VAL?
enter statement
date of April 4 and next cal due date as
October 4 (YY = year of due date).
queries the calibration message and returns a quoted string
Query the calibration message.
Enter value into computer.
>
specifies the value of the known calibration
Parameter Type
Range of Values
numeric
See the service manual
Enter calibration value.
Query the calibration value.
Enter value into computer.
Default Units
none