*OPC?

Comments

*RST

Comments

*SRE and *SRE?

Parameters
Chapter 3
*OPC?
causes the E1412A to wait for all pending operations to complete. A single
ASCII "1" is then placed in the output queue.
Executable when Initiated: Yes
Coupled Command: No
Related Commands:
*RST
Condition: none
*RST
resets the E1412A as follows:
-- Sets all commands to their
-- Aborts all pending operations.
does not affect:
*RST
-- The output queue
-- The Service Request and Standard Event Status Enable Registers
-- The enable unmasks for the Questionable Signal Registers
-- Calibration data
Executable when Initiated: Yes
Coupled Command: No
Condition: none
*RST
unmask
specifies which bits of the Status Byte Register are enabled
*SRE <
>
(unmasked) to generate a IEEE-488.1 service request. Event and summary bits are
always set and cleared in the Status Byte Register regardless of the unmask value.
<
unmask
>
is the sum of the decimal weights of the bits to be enabled allowing these
bits to pass through to the summary bit RQS (bit 6 in the status byte).
*SRE?
returns the current enable unmask value.
Parameter Name
Parameter Type
unmask
<
>
A 1 in a bit position enables service request generation when the corresponding
Status Byte Register bit is set; a 0 disables it.
*OPC
,
*WAI
*RST
state.
Range of Values
numeric
0 through 255
Multimeter Command Reference
Default Units
none
165