The Trigger Source

46 Agilent E1412A Multimeter Application Information
Triggering the multimeter is a multi-step process that offers triggering
flexibility.
1. You must configure the multimeter for the measurement by selecting
the function, range, resolution, etc.
2. You must specify the source from which the multimeter will accept
the trigger. The multimeter will accept a BUS trigger from the
VXIbus, an external trigger from the front panel "Trig" BNC
connector or an immediate trigger from the multimeter's internal
trigger system.
3. You must make sure that the multimeter is ready to accept a trigger
from the specified trigger source (this is called the wait-for-trigger
state) by issuing a
always uses an immediate trigger (see the flow chart in Figure 2-1 on
page 45).
The
TRIGger:SOURce <
trigger system to respond to the specified source. The following trigger
sources are available:
BUS:
Trigger source is the GPIB Group Execute Trigger (GET) or the
common command. Within the 75000 Series C mainframes, the
*TRG
instrument whose trigger source is "
addressed to listen will respond to the GPIB Group Execute Trigger.
The
command differs from GET because it is sent to a specific
*TRG
instrument not a group of instruments. NOTE: B-size controllers do
not support the BUS trigger (e.g., E1306A command module,
E1300/E1301A B-size mainframes).
EXTernal:
Trigger source is the multimeter's external trigger BNC
connector (labeled "Trig" on the front panel). A falling (negative-going)
edge of the input signal triggers the multimeter. The external pulse
signal must be >1s, +5V maximum to 0V (TTL levels).
IMMediate:
Internal trigger is always present. If the multimeter is in
the wait-for-trigger state (
sends the trigger. The
automatically set the trigger source to
TTLTrg0
through
TTLTrg7:
lines. The multimeter is triggered on the falling (negative- going) edge
of a TTL input signal. NOTE: B-size controllers do not support
VXIbus TTL triggers (e.g., E1306A Command Module,
E1300/E1301B B-Size Mainframes).
For example, the following program statement selects the external trigger
BNC connector as the trigger source.
TRIGger:SOURce EXTernal
You can change the trigger source only when the multimeter is in the idle
state. Attempting to change the trigger source while the multimeter is in the
wait-for-trigger state will generate the "Settings conflict" error.
READ?
or
INIT
command. A
source
command configures the multimeter's
>
" and was the last instrument
BUS
),
INITiate
TRIGger:SOURce IMMediate
and
MEASure
CONFigure
IMMediate
Trigger source is the VXIbus TTL trigger
MEASure
command
commands
.
Chapter 2