:CURRent:AC?

Parameters
Comments
Example
102 Multimeter Command Reference
MEASure:CURRent:AC? [<
resolution
[,<
>|MIN|MAX|DEF]
specify the measurement range and resolution (see range versus resolution table at
start of chapter).
Parameter Name
Parameter Type
range
<
>
resolution
<
>
To select a standard measurement range, specify range as the input signal's
maximum expected current. The multimeter then selects the correct range that
will accept the input.
The
or
AUTO
DEFault
The
and
MIN
MAX
range and resolution:
For range:
= 1A;
MIN
For resolution: see Table 3-5 on page 71 for resolution choices.
To select autorange, specify
parameter. In the autorange mode, the multimeter samples the input signal
before each measurement and selects the appropriate range.
To specify a
or
MIN
AUTO
or
DEFault
parameter for the range parameter as
MEAS:CURR:AC? DEF
the
or
resolution from being interpreted as a range setting and the
MIN
MAX
resulting command becomes
MEAS:CURR:AC? DEF,MAX
Related Commands:
Making AC Current Measurements
MEAS:CURR:AC? 1,MAX
range
>|MIN|MAX|DEF|AUTO
] selects the AC current function and allows you to
Range of Values
numeric
MIN|MAX|DEF|AUTO
numeric
| MIN | MAX | DEF
option for the range parameter enables autorange.
parameters select the minimum or maximum values for
= 3A
MAX
DEF
for range or do not specify a value for the
resolution while autoranging, you must specify the
MAX
(you cannot omit the range parameter). This prevents
MEAS:CURR:AC? DEF,MIN
(or use
AUTO
FETCh?
,
INITiate
,
READ?
Function: AC Current; range selected: 1A;
MAX resolution: 1.0E-04 A.
Default Units
1A|3A|
resolution|
or
in place of
DEF
).
A
A
Chapter 3