:LFRequency?

Comments
Example

:SECure:CODE

Comments
Example
80 Multimeter Command Reference
CALibration:LFRequency?
This command returns
+50
multiple of 50.
Query the Line Frequency Setting
CAL:LFR?
CALibration:SECure:CODE <
To change the security code, first unsecure the multimeter using the old security code
with
CAL:SEC:STAT OFF, <
security code may contain up to 12 characters. The security code is stored in
non-volatile memory.
The security code is set to "HP_
shipped from the factory. The security code is stored in non-volatile memory,
and does not change when power has been off or after a remote interface reset.
The security code
<
The first character must be a letter. The remaining characters can be letters or
numbers or an underscore. You do not have to use all 12 characters but the first
character must be a letter.
If you forget or lose the active security code, you can disable the security
feature by adding a jumper inside the multimeter (see Chapter 5 in the Service
Manual). You then enter a new code and remove the jumper.
Enter a New Calibration Security Code
CAL:SEC:STAT OFF, HP_E1412
the_new_code
CAL:SEC:CODE
queries the line frequency setting.
for line frequency set to 400 because 400 is an even
Query the line frequency.
new code
enters a new calibration security code.
>
old code
>
. Then, enter the new code. The calibration
E1412
" for C-size when the multimeter is
new code
can contain up to 12 alphanumeric characters.
>
Unsecure with the old code.
Enter a new calibration code
(a maximum of 12 characters).