RESistance:RESolution

Parameters
Comments
Example
RESistance:RESolution?
Chapter 3
[SENSe:]RESistance:RESolution <
resistance measurements.
Parameter Name
Parameter Type
resolution
<
>
selects the best resolution (the smallest value) for the selected range.
MINimum
selects the worst resolution (the largest value) for the selected range.
MAXimum
You must select a range using
resolution. Also, only specify a resolution when making measurements on a
fixed range. Otherwise, the resolution will change to correspond with the range
selected during autoranging.
If autoranging is required, set the resolution using the
If necessary to achieve the specified resolution, the multimeter will increase
the integration time as needed. This command overrides the results of
previously executed
(the last command executed has priority).
The
RES:RESolution
previous
CONFigure:RESistanc
Related Commands:
Condition: Based on the
*RST
command.
Change the Resolution
CONF:RES 1560,MAX
RES:RES 10E-03
READ?
enter statement
[SENSe:]RESistance:RESolution? [MIN|MAX]
numbers to the output buffer.
The present resolution selected if
resolution values available on ranges set by the
returned.
The resolution with the smallest value (i.e., the best resolution) for the selected
range if
MIN
is specified.
The resolution with the largest value (i.e., the worst resolution) for the selected
range if
is specified.
MAX
resolution
>
Range of Values
numeric
resolution
RESistance:RANGe
RESistance:NPLC
or
RESistance:APERture
command overrides the resolution setting from a
e command.
e,
CONFigur
RESistance:NPLC
values for the
*RST
Function: 2-wire ohms; range selected: 10k;
MAX resolution: 1.
Set resolution to 10m.
Place multimeter in wait-for-trigger state and make
measurements; send readings to output buffer.
Enter readings into computer.
or
MIN
MAX
Multimeter Command Reference
selects the resolution for 2-wire
Default Units
ohms
|MIN|MAX
before specifying
or
parameters.
MIN
MAX
command
RESistance:NPLC
returns one of the following
are not specified. Only the
command are
RANGe