RESistance:RESolution
Parameters
Comments
Example
RESistance:RESolution?
Chapter 3
[SENSe:]RESistance:RESolution <
resistance measurements.
Parameter Name
Parameter Type
resolution
<
>
•
selects the best resolution (the smallest value) for the selected range.
MINimum
selects the worst resolution (the largest value) for the selected range.
MAXimum
•
You must select a range using
resolution. Also, only specify a resolution when making measurements on a
fixed range. Otherwise, the resolution will change to correspond with the range
selected during autoranging.
•
If autoranging is required, set the resolution using the
•
If necessary to achieve the specified resolution, the multimeter will increase
the integration time as needed. This command overrides the results of
previously executed
(the last command executed has priority).
•
The
RES:RESolution
previous
CONFigure:RESistanc
•
Related Commands:
•
Condition: Based on the
*RST
command.
Change the Resolution
CONF:RES 1560,MAX
RES:RES 10E-03
READ?
enter statement
[SENSe:]RESistance:RESolution? [MIN|MAX]
numbers to the output buffer.
•
The present resolution selected if
resolution values available on ranges set by the
returned.
•
The resolution with the smallest value (i.e., the best resolution) for the selected
range if
MIN
is specified.
•
The resolution with the largest value (i.e., the worst resolution) for the selected
range if
is specified.
MAX
resolution
>
Range of Values
numeric
resolution
RESistance:RANGe
RESistance:NPLC
or
RESistance:APERture
command overrides the resolution setting from a
e command.
e,
CONFigur
RESistance:NPLC
values for the
*RST
Function: 2-wire ohms; range selected: 10k;
MAX resolution: 1.
Set resolution to 10m.
Place multimeter in wait-for-trigger state and make
measurements; send readings to output buffer.
Enter readings into computer.
or
MIN
MAX
Multimeter Command Reference
selects the resolution for 2-wire
Default Units
ohms
|MIN|MAX
before specifying
or
parameters.
MIN
MAX
command
RESistance:NPLC
returns one of the following
are not specified. Only the
command are
RANGe