Service
Verifying Oscilloscope Performance
To verify bandwidth (alternate method)
In this test you verify the bandwidth of the oscilloscope by using a power
meter and power sensor to set the output of a signal generator at 1 MHz and
the upper bandwidth limit. You use the peak-to-peak voltage at 1 MHz and
the upper bandwidth limit to calculate the bandwidth response of the
oscilloscope.
Test limits
HP 54600B and HP 54601B, all channels (−3 dB)
HP 54602B
Channels 1 & 2 (–3 dB) *
Channels 3 & 4 (–3 dB)
HP 54603B, all channels (−3 dB)
Table 4-7
Equipment Required
Equipment
Signal generator
Power meter and
Power Sensor
Power splitter
Cable
Adapter
Feedthrough
*
1mv/div to 5 mv/div dc = 100 MHz.
4–12
dc to 100 MHz
ac coupled 10 Hz to 100 MHz.
dc to 150 MHz
ac coupled 10 Hz to 150 MHz.
dc to 250 MHz.
dc to 60 MHz
ac coupled 10 Hz to 60 MHz.
Critical specifications
1 to 100 MHz at 200 mV
(250 MHz for HP 54602B)
1 to 100 MHz ±3% accuracy
(250 MHz for HP 54602B)
Outputs differ by <0.15 dB
Type N (m), 24 inch
Type N (m) to BNC (m)
50 Ω, BNC (m) and (f)
Recommended Model/Part
HP 8656B opt 001
HP 436A and HP 8482A
HP 11667B
HP 11500B
HP 1251-0082
HP 10100C