HP 54600B User, Installation And Service Manual page 97

Hide thumbs Also See for 54600B:
Table of Contents

Advertisement

Service
Verifying Oscilloscope Performance
Set up the oscilloscope.
1
, then press the Default Setup softkey.
Press
a
Setup
, then press the V avg softkey.
Press
b
Voltage
Set the Volts/Div to the first line of table 4-4.
c
Adjust the channel 1 Position knob to place the baseline near
d
(but not at) the bottom of the display.
Press
, then press the
2
Cursors
Using the cursors knob, set the V1 cursor on the baseline.
3
If you are in an electrically noisy environment, it can help to place a shorting
cap on the input BNC connector when positioning V1.
Connect the power supply to the oscilloscope and to the multimeter,
4
using the BNC tee and cables.
Set the power supply output to the first line in table 4-4.
5
4–8
softkey.
V1