To verify voltage measurement accuracy

In this test you verify the voltage measurement accuracy by measuring the
output of a power supply using dual cursors on the oscilloscope, and
comparing the results with a multimeter.
Test limits ±1.9% of full scale (HP 54600B, HP 54601B, HP 54602B)
Table 4-3
Equipment Required
Equipment
Power supply
Digital multimeter
Cable
Shorting cap
Adapter
Adapter
±2.4% of full scale (HP 54603B)
Critical specifications
14 mV to 35 Vdc, 0.1 mV resolution
Better than 0.1% accuracy
BNC, Qty 2
BNC
BNC (f) to banana (m)
BNC tee (m) (f) (f)
Verifying Oscilloscope Performance
Recommended
Model/Part
HP 6114A
HP 34401A
HP 10503A
HP 1250-0774
HP 1251-2277
HP 1250-0781
Service