General - HP 54600B User, Installation And Service Manual

Hide thumbs Also See for 54600B:
Table of Contents

Advertisement

General

Environmental
The instrument meets or exceeds the environmental requirements of
characteristics
MIL-T-28800D for Type III, Class 3, Style D equipment as described below.
Ambient temperature (Tested to MIL-T-28800D paragraphs 4.5.5.13
option 2 and 4.5.5.14)
Operating –10 °C to +55 °C
Nonoperating –51 °C to +71 °C
Humidity (tested to Hewlett-Packard environmental specification section
758 paragraphs 4.0, 4.1, and 4.2 for class B-1 products)
Operating 95% relative humidity at +40 °C for 24 hours
Nonoperating 90% relative humidity at +65 °C for 24 hours
Altitude (Tested to MIL-T-28800E paragraph 4.5.5.2)
Operating to 4,500 m
Nonoperating to 15,000 m
EMI
EMI (commercial) CISPR 11 (ISM, Group 1, class A equipment)
EMI Meets the requirements in accordance with MIL-T-28800D,
CE01 Part 2 narrow band requirements up to 15 kHz
CE03 Part 4
CS01 Part 2
CS02 Part 2
CS06 Part 5 limited to 300 V